ELIT Brand Electrochemical Sensors and Computer-based Instrumentation.
Innovative technology at highly competitive prices.
HOME PRODUCTS TECHNICAL INFORMATION SITE MAP CONTACT US PRICE LIST

ELIT Electrochemical Signal Converters

Enable the use of electrochemical sensors in industrial and laboratory applications using any type of data acquisition or process-control systems.

  • Use sensor signals to activate alarms or process-control equipment at critical values.
  • Rugged, dust and splash-proof units - ABS enclosure, 200mm x 120mm x 75mm,
        IP65 Protection Class.
  • Wide input range power supply: 18 to 36 V DC, 150 mA
  • Sensors galvanically isolated from Power supply and Output.
  • Maximum flexibility and environmental protection at minimum cost.

    Please e-mail nico2000ltd@btinternet.com with details of your requirements
    and we will advise on which version is most appropriate for your application.


  • SC1-ISE1-A Converter (1-Channel ISE, 4 to 20 mA, 0 to 5V)

  • Input: pH-, ORP-, Ion-selective Electrode via BNC connector.
  • Output: 4-20 mA via 2 PCB Terminals, Cable Inlet for cables with 4 to 10 mm Diameter via IP 65 Cable Gland.
  • Power Supply: Wide-range 24 V DC power supply. 18V DC to 36 V DC are accepted. 200 mA power consumption.
  • Isolation: The input is fully isolated from the Output and the Power Supply.
  • Enclosure: Cream colour ABS, dimensions: 200 x 120 x 75 mm, IP 65, with opaque or transparent lid, with fixture holes isolated from the compartment of the enclosure.
  • Transfer Function: A range of 15 gain-settings and 15 Offset-settings together with Gain-polarity and Offset-polarity enables adjustment to any Electrode Input Range between -1000 mV and +1000 mV. This "Zoom-function" makes precise Ion-selective measurements possible.
  • Digital Display: A built-in Digital Display simplifies precise adjustment and Calibration.


    CCR - Nico2000 Ltd, London, UK - Last Update 19 Jan 2012