ELIT Brand Electrochemical Sensors and Computer-based Instrumentation.
Innovative technology at highly competitive prices.
HOME PRODUCTS TECHNICAL INFORMATION SITE MAP CONTACT US PRICE LIST

Method for determining the concentration of AMMONIUM (NH4+)
in aqueous solutions.

Electrode Specifications           Links to publications on the WWW         Go to Ion Selective Electrodes page

Many published ISE methods for ammonium use an ammonia gas-sensing electrode to measure the NH3 gas liberated after the sample has been treated with strong alkali. The ELIT ammonium electrode measures the NH4+ ion directly in solution and there is no need for any pre-treatment with caustic chemicals, and no noxious gases are liberated.

Note 1: Because of the interference of K+ on the NH4+ electrode, there is a special method for measuring samples with a K content greater than half that of the NH4

Note 2: In aqueous solution, there is an equilibrium between NH4+ and NH3 which is pH dependent.
Below about pH 6.5 there is 100% NH4+; above about pH 11.5 is 100% NH3
Thus, if TAN (Total Ammonia Nitrogen) is required then the pH of the samples must be kept below 7 (Note that pure NH4 standard solutions have a pH of about 5.5 so should always be fully ionised). However, if the level of dissolved NH3 (toxicity) is more important then the pH and temperature of each sample must also be measured and the NH4/NH3 ratio calculated.

For more information, see:
Relationship between NH4/NH3/pH/Temp. in aqueous solutions.
Ionization and Toxicity of Ammonia.
Estimation of the Allowable Upper Limit of Ammonia in Saline Waters.
Detailed Investigation of various factors to consider when making NH4 measurements.

Apparatus Required:

Ion-Selective Electrode for ammonium ion (ELIT 8051 PVC membrane).

Reference electrode: double junction lithium acetate (ELIT 003n).

Dual electrode head (ELIT 201).

ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter.

Standard solution: 1000 ppm NH4 as NH4Cl.

Buffer solution (ISAB): 1 Molar CuSO4.

150 ml polypropylene beakers, 100ml volumetric flask, 1, 2, 5, 10ml pipettes.


Standard Solution Preparation

For the most accurate results, Ammonium chloride must be dried for one hour at 100 °C. and cooled in a dessicator. To make 1000ppm NH4+, dissolve 2.965 g of NH4Cl in deionised water in a 1000 ml volumetric flask and dilute to the mark. According some sources, this solution is stable for six months when stored in a refrigerator but more dilute solutions for preparing the calibration curve should be freshly made each day. However, tests made at Nico2000 Ltd have demonstrated that a 10,000ppm solution, stored on the laboratory shelf at room temperature, remained stable and accurate for more than 8 years, 100ppm was still accurate after 10 days standing in a covered beaker, and 10 & 1 ppm were OK after 4 days.

Calibration:

Before use, the electrodes must be calibrated by measuring a series of known standard solutions, made by serial dilution of the 1000ppm standard solution. For a full calibration, prepare 100ml of solutions containing 1000, 100, 10, 1, and 0.1ppm NH4. If the approximate range of concentration of the samples is known, and this is within the linear range of the ISE, then it is only necessary to make two solutions (preferably a decade apart) which span this range.

Follow the instructions on the General Operating Instructions to measure these standard solutions and prepare a calibration graph.

NB: If the samples to be measured are expected to have a total ionic strength of greater than 0.01 Molar, then add 10 ml of buffer solution to each standard and mix thoroughly to compensate for different activity coefficients between samples and standards. BUT note that this will change the calibration slightly (ie: increase the slope but also raise the lower limit of linearity and detection limits) and lower the pH so that any NH3 will be converted to NH4.
See Typical calibration graphs (Excel file)

Sample Preparation:

For low ionic strength samples, no sample preparation is necessary. Simply take approximately 50 to 100 mls of sample in a plastic beaker, or even immerse the electrodes directly in a lake or river (but take care to avoid losing the electrodes!). For high ionic strength samples, take 100mls of sample and add 10 mls of buffer solution and stir well before measurement.

Sample Measurement:

Follow the instructions in the electrode operating instructions to measure a series of samples and record the results. Briefly, it is important to note that, if measuring in beakers, the electrodes must be washed and dried between each sample, to avoid cross contamination, and sufficient time must be allowed (2 or 3 minutes), before taking a reading after immersion, to permit the electrode signal to reach a stable value. For the highest precision, frequent recalibration is recommended (see operating instructions).

Results:

The results will be displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples then these figures will not need adjusting because they will all be affected by the same dilution factor.


Return to top

Procedure for determining the concentration of AMMONIUM (NH4+) in aqueous solutions with high potassium content.

For samples with K/NH4 ratios greater than about 0.5, rather than relying on the general Selectivity Coefficient quoted in the electrode specifications to assess the likely level of interference, a more reliable correction can be made by measuring the interference directly in a typical sample. i.e. first measure the NH4 and K in the sample, then add more K (sufficient to ensure a significant increase in signal - eg double the concentration) and measure the apparent NH4 concentration again. Then calculate the SC from the increase in NH4 caused by known addition of K. However, it is recommended that the accuracy and precision of this correction should be evaluated for each particular application before too much reliance is placed on it. This measurement can be made, with or without ISAB, by Direct Potentiometry or Standard Addition depending on the Ionic Strength of the samples.

A detailed description of the procedure can be found in: Examples of measurement of rain, pond, and river water with correction for K.

Alternatively, if the level of K is constant in the samples, then a compensation can be made by spiking the standards with the same concentration of K so that the standards have the same level of interference as the samples.

Return to top

Technical Specifications
for the ELIT Ammonium Ion-Selective Electrode (ELIT 8051)

Click here to download a printer-friendly (pdf) Specification Sheet.

Introduction
The Ammonium Ion-Selective Electrode has a solid-state PVC polymer matrix membrane which is designed for the detection of ammonium ions (NH4+ ) in aqueous solutions and is suitable for both field and laboratory applications.
The Ammonium Ion is a monovalent cation .
One mole of ( NH4+ ) has a mass of 18.038 grams; 1000 ppm is 0.055 M
Dissolve 2.965g anhydrous ammonium chloride (NH4Cl) in 1 Litre water.

Physical Specifications
Length of body excluding gold contact = 130 mm
Length of body including gold contact = 140 mm
Diameter of body = 8 mm
DC resistance at 25°C = 2.5 MOhm
Minimum feasible sample volume = 5 ml

Chemical / Operational Specifications
Preconditioning / Standard solution : Normally 1000 ppm NH4+ as NH4Cl
(But see General Operating Instructions)
Preconditioning time : at least 5 minutes
Optimum pH range : pH 0 to pH 8.5
Note: this is the range in which the electrode response to NH4+ is not affected by changes in pH.
For accurate measurements of total nitrogen, pH should be kept below 7 to ensure 100% ionisation.
Temperature range : 0 to 50° C
Recommended ISAB : 1M MgSO4 (OR 1M CuSO4 for NH4 < 50ppm) (Add 10% v/v)
Recommended reference electrode : double junction (ELIT 003)
Reference electrode outer filling solution : 0.1M CH3COOLi.
Electrode slope at 25°C : 54 ± 5 mV/ decade
Concentration range : 0.03 to 1,800 ppm (2x10-6 to 0.1 Molar)
Response time : 10 seconds
(Defined as time to complete 90% of the change in potential after immersion in the new solution.)
Time for stable reading after immersion : < 1 to > 5 minutes
(Depending on concentration, use of ISAB, nature of sample and stabilisation time of liquid junction potential of reference electrode.)
Potential drift (in 1000 ppm) : < 3 mV/ day (8 hours)
(Measured at constant temperature and with ISE and Reference Electrode continually immersed.)

Interference:

The following ions interfere with the ammonium measurement - Selectivity Coefficients (SC)  in brackets: Potassium (0.1), Sodium (0.002), Magnesium (0.0002), Calcium (0.00006), Lithium (0.00003).

The SC is the approximate apparent  increase in the measured concentration caused by 1 unit of the interferent.  Thus the likely effect of any interfering ion (% increase) can be calculated as follows:  ((expected concentration) x (SC) / (expected NH4 concentration)) x 100.

On this basis, only the K interference is likely to cause a significant error.

However it must be noted that if MgSO4 is used as ISAB it can cause interference on low-level samples. Direct measurements have shown that the SC for 0.1M Mg can be considerably higher than that quoted by most manufacturers (and can give an apparent NH4 signal of up to 2ppm) so it is recommended that CuSO4 should be used instead for samples with less than about 50 ppm NH4.

Return to top

Last update: CCR, 25 Nov. 2016 .